Sonatest supply a wide variety of Ultrasonic and other Non destructive testing instrumentation


Alpha Gage - A - Scan Corrosion Gage

The Alphagage includes these features:

  • Storage for up to 8000
    thickness readings
  • Unique Help Function
  • Resolution 0.01 mm
  • Through paint measurement
  • Compatible with any standard
    flaw detection probe
  • Suitable for harsh environment (IP67 rated)
  • infrared data port for PCS
    and peripherals
  • Winsat Windows 95 Software
    for PC interfacing

Alphagage Specification:

Display: 18 x 71mm (1.9 x 2.8 in) viewable area, variable contrast setting.
A-scan: 128 x 96 pixels, outline or filled trace, Auto freeze on loss of signal, Positive, negative and full wave rectification.
Update rate: 5 to 10 Hz.
Measurement Resolution:

0.01 mm (0.001 in)

Measurement modes: Depth to first echo, Echo - to - Echo, Minimun stackness mode, V path correction for dual probes.
Minimum thickness: Less than (0.6 mm to 0.0237 in) depends on transducers
Transmitter: 20, 60, and 200 volt, variable pulse width 50, 100, 250 ns.
TCG: 80 db dynamic range, maximum 6db per microsecond.
Range: 5mm to 500 mm (0.25 in to 20 in)
Velocity: 2000 m/s to 9999 m/s (0.0787 to 0.3937 in/µs
Probe Zero; 0 to 10000 microseconds
Gain: 0 to 80 db in 0.3 db steps
Frequency:

1 to 7.5 MHz 5 db, 0.8 to 9.5 Mhz -6db.

Calibration modes: Zero block, two point autocalibration, Manual input of Zero and Velocity.
Thickness Logging; Storage for 2500 readings with A-scan, 8000 readings with thickness only, Recall of thickness and A-scan, 48 character alphanumeric thickness identification, Historical phase and thickness reading.
Panel Storage: Storage for 100 setups.
Serial Communication: Infrared communication to printer and PC.
Battery: NiCd LAH 9 hours battery duration, Auto shutdown to preserve battery life.
Enclosure: Sealed to IP67.
Operating temprature range: 40 to 55 degrees celsius.

Size:

285 x 120 mm Maximum (11.2 x 4.7 in)
Weight: 0.88 kg (1.94 lbs) includes battery.
Charger: 90 to 260 v AC, 47 - 440 Hz 2 hour test charge, Discharge feature for battery conditioning, Operating temperature 10 to 30 degrees.
Probes: Up to 10 pre-defined Sonatest Products, 5 user configurable, Dual, stage or delay probes available.
Keypad: Sealed keypad with tactile and audiable response.
Alarm: Minimum thickness alarm.
Operator ID: 16 charaters
Freeze Mode: Option to freeze display.
Help: On line help facility.
Accessories & Transducers: Leather case, Spare battery, SGI, SG3, SG2HT, SLB5-10, GEM5-10SF
Winsat Software: Thickness logging for PC interfacing and reading analysis Windows licensed software, Fully bidirectional transfer of data between PC and Alphagage.