C75 On-site Multifunction Calibrator
Model Number C75
Product
Specifications:
- Temperature
Resistive Probes:
Pt 50 to 1,000; Ni 100 to 1,000; Cu 10 to 50
Example: Pt 100:
Accuracy/one year Measurement = 0.012%R+0.05°C,
0.01°C Resolution
Accuracy/one year Emission = 0.014%R+0.12°C,
0.02°C Resolution
- Thermocouples:
13 different types, K, T, J, E, R, S, B, U, L, C, N, Platine,Mo
Example: Type K
Accuracy/one year Measurement & Emission =
0/1372C = 0.013%R+0.08°C, 0.05°C Resolution
- DC
Current:
Example: up to 50mA with 1µA Resolution
Accuracy/one year measurement & emission = 0.018%R+2µA
- DC
Voltage:
Four ranges, +/-100mV, 1V, 10V, 50V
Example: +/-100mV with 1µV Resolution
Accuracy/one year measurement & emission = 0.013%R+3µV
(+2µV)
- Resistance:
Two ranges, 400ohms, 4000ohms
Example: 400ohms with 1m ohms resolution
Accuracy/one year 0.012%R+10m ohms Measurement
- Frequency
and Counting:
Example: 20KHz with 0.01Hz
Accuracy/one year = 0.005%R
|

|