FORMSCAN 3000 SPECIFICATIONS

Rotary Table/Spindle Accuracy:

 

Radial departure from least squares circle (at 1"/25mm above table) - not to exceed ±1.0u"/±0,025um
Axial Deviation - not to exceed 1.0u"/ 0, 025 um
Coning - not to exceed 0.5u"/inch; 0,5um/m
Total System Accuracy:  


(Rotary Table/Spindle, Amplifier/ Recorder, Gage Heads and Auto-Centering Computer)

±1% of full scale
or 2u"/0,05um, whichever is larger


Operating Range Minimum Graduation Chart Magnification ("MAG. x 1 ")
±.0001 "/±2 um
±.0002"/ ±4 um
±.0004"/ ± 10 um
±.001 "/±20um
±.010"/ ±200 um
5u"/0,1 um
10 u"/ 0, 2 um
20 "/0,5um
50 u"/ 1, 0 um
500u"/ 10,0um
20000 X /25400 X
10000X/ 12700 X
5000X/5080 X
2000X/2540X
200 X /254 X

Rotary Table/Spindle:  
Table Diameter Rotational Speed 8"/ 200 mm
Rotational Speed 4 rpm (fixed)
Radial Flatness (concave) - not to exceed 100u"/ 2,5um
Circular Flatness - not to exceed 25 u"/ 0,63 um
Maximum Support Capability 200lb/90kg (balanced load)
Spindle Stiffness- Axial Spindle 1.5x 1E6 lbf/inch; 2,6x1E8 N/m
Stiffness- Moment 70 in Ib/arc sec; 7,9 N - m/arcsec
Centering Adjustment Tilt/Centering Table AT-96
Tilt Adjustment (optional accessory)
Maximum Workpiece Radius 91/4"/ 235 mm
Maximum Gaging Height 16"/ 406 mm
Gage Head Arm Adjustment 8"/ 200 mm
Center Hole 11/2"/38mm dia., 25/8"/65mm deep
Support Pressure 50psig; 3,5kgf/cm2
Air Consumption (max.) 1.5scfm; 2,6m3/hour
Source Pressure (min.-max.) 80-125psig; 5,6-8,75kgf/cm2
Air Filter/ Regulator Adjustment 0 to 100 psig and 0 to 7 kgf/cm2

Amplifier:  
Repetition Accuracy To within 2u"/0,05um or 1/10 scale division, whichever is larger
Calibration Accuracy To within 4/5 scale division
Setup Meter Analog, ±20 graduations
Range of Zero Adjustment ±.002"/±50 um
Stability - Short Term Less than 0.01 % change of gain over an 8 hour period
Stability- Long Term Less than 0.1 % change of gain over a 1 month period
Thermal Drift 0.01 % per °F/ 0,018% per °C
Power  
- Inch/Metric Models 50 watts, 117 Vac ± 10%, 50-60 Hz
or 25 watts, 220 V ac ±10%, 50-60 Hz
- Metric only Models 25 watts, 220Vac ±10%, 50-60 Hz
Signal Output ±10Vdc at 5mA; 75 Ohm effective output impedance. Output is short-circuit proof; short-circuit current is 25mA

Recorder:  
Polar Chart Range ±10 graduations
Polar Chart Diameter 6 "/ 150 mm
Rotational Speed 4 rpm (fixed)
Recorder Pen Heat writing (thermal)
Recorder Response Settings 0-15, 0-50, 0-150 and 0-500 undulations/revolution

Gage Head: Lever type, Model EHE-1056, with adjustable nose mount
Contact Point 062"/ 1,6mm diameter steel ball tip, EPT-1004
Measuring Range ±.010"/±0,25 mm
Pretravel and Overtravel 005"/ 0,13 mm
Gaging Force approximately 2,5g
Range of Fine Adjustment 150"/ 3,8 mm

Auto-Centering Computer:  
Resolution 0.1 division
Power  
- Inch/Metric Models 200 watts, 117 V ac ± 10%, 50-60 Hz
or 100 watts, 220 V ac ± 10%, 50-60 Hz
- Metric only Models 100 watts, 220 V ac +/-10%, 50-60 Hz

Click here for the latest preowned deals, sold as is, where is. No returns or refunds.

Contact Us for additional information



Prepaid wire transfer required for all international orders
 [Home] [About Us] [Products] [Services] [What's New] [Tech. Info] [Contact Us] [Leasing]
All Copyrights reserved by Deterco, Inc. 2007
Tel: (713) 943-2721 e-mail: sales@deterco.com